ACT Option
The ACT option supports four types of electrical transport current measurements: resistivity, hall coefficient, I-V curve and critical current. Measurements are typically made by passing an AC excitation current through the sample and measuring the AC voltage drop across the sample.
The ACT option is compatible with the He-3 system..
Puck estándar (2 muestras)
Standard puck
10.0 mm x 10.0 mm
(2 samples)
  • Temperature Range: 1.9 K a 400 K.
  • Magnetic field range: 0 a ± 9 T, ± 14 T. *
  • Current: ± 0.02 μA — 2 A.
  • Frequency: 1 Hz – 1 k Hz.
Puck He-3 (1 muestra)
Puck He-3
10.2 mm x 11.2 mm
(1 sample)
  • Temperature Range: 350 mK a 350 K.
  • Magnetic field range: 0 a ± 9 T, ± 14 T. **
  • Current: ± 0.02 μA — 20 mA.
  • Frequency: 1 Hz – 1 k Hz.
  • Resistivity, I-V curve and critical current measurements:
    • Four-wire measurementss
      Medida típica por 4 puntos
    • Resistivity: AC Mode: an AC excitation current is passed through the sample and the in-phase component of the AC potential drop across the sample is reported. The frequency filtering of the output signal provides better sensitivity than DC resistivity measurements.
    • I-V curve: a DC current IDC through the sample is ramped while the voltage drop is measured. I-V plots may be especially interesting when nonlinear behavior exists.
    • Critical current measurements. As an ohm meter, the ACT option provides a convenient method of determining the critical current of superconducting samples.
  • Hall coefficient measurements:
    • Five-wire measurements
    • A five-wire configuration and a compensation technique allows measuring only the potential difference due to the Hall potential ***
      Medida típica por 5 puntos
    • AC Mode: an AC excitation current is passed through the sample and the in-phase AC potential drop across the sample is measured. The frequency filtering of the output signal provides better sensitivity than DC resistivity measurements.
    • An external junction box is available to reconfigure measurement channels assignation.
    * The magnetic field is applied perpendicular to standard puck.
  ** He-3 option allows sample puck orientation either parallel or perpendicular to the magnetic field.
*** Sample resistance must be lower than potentiometer resistance ~100 Ω.